Search productSearch post

Air Fluoride Sampling Device
Air Fluoride Sampling Device
Air Fluoride Sampling Device
Air Fluoride Sampling Device

Air Fluoride Sampling Device

  • Keywords:air fluoride sampling device, fluoride measurement, airborne fluoride detection, environmental monitoring
  • Category:CO2 Analysis and Detection

The Air Fluoride Sampling Device is a specialized instrument designed for the accurate collection and analysis of fluoride compounds in the air. This device utilizes advanced sampling technology to provide precise measurements of airborne fluoride concentrations, ensuring compliance with environmental and safety standards. It is ideal for use in industrial settings, environmental monitoring, and research applications. The robust design and high sensitivity make it suitable for detecting even low levels of fluoride in the atmosphere.

CLICK TO INQUIRE

Product Overview
The Air Fluoride Sampling Device is a specialized equipment designed for collecting fluorides in the air. It uses filter membrane sampling or fluoride ion-selective electrode methods, offering stable performance and ease of operation. The device is versatile and can be used not only for fluoride sampling but also for simultaneously collecting other pollutants such as heavy metals, TSP, PM10, PM2.5, and aerosols in the ambient air. It is widely applied in environmental protection, public health, labor safety, industrial monitoring, scientific research, and education, especially for the collection and study of fluorides, heavy metals, and SVOCs.

Key Features

  • Compliance with National Standards: The device complies with the "Environmental Air Fluoride Determination using Filter Membrane Sampling/Fluoride Ion Selective Electrode Method" (HJ 955-2018) standard, with a wide sampling flow range (16.7-110L/min).
  • High Negative Pressure Pump: Equipped with a high-negative pressure pump, the device can achieve a maximum flow rate of 50L/min, overcoming resistance above 22kPa, making it suitable for heavy metal and particulate matter sampling.
  • Automatic Recovery Function: In the event of a power failure during sampling, the device automatically protects data and restores the sampling process once power is restored, ensuring data integrity.
  • Data Export Function: Supports USB data export and Bluetooth printing, facilitating data storage and report generation for analysis and documentation.

Applications

  • Environmental Monitoring: Ideal for monitoring environmental air pollutants such as fluorides, heavy metals, and particulate matter.
  • Scientific Research: Provides accurate sampling data for environmental science and air quality research.
  • Labor Safety: Used to monitor harmful substances like fluorides in the workplace, ensuring worker health and safety.
  • Emergency Monitoring: Serves as a rapid-response tool during environmental pollution incidents.
  • Education and Training: Suitable for educational and experimental purposes in environmental monitoring technology.
Main Parameter Range Resolution Accuracy
Sampling Flow (16.7~110) L/min 0.1L/min Better than ±2.5%
Flow Stability / / Better than ±2.0%
Flow Repeatability / / Better than ±2.0%
Pre-pressure (-30~35) kPa 0.1kPa Better than ±2.5%
Atmospheric Pressure (60~120) kPa 0.1kPa Better than ±0.5kPa
Noise <68dB(A)
Power Supply AC220V±10% 50Hz
Power Consumption <150W
Overall Dimensions 280mm×270mm×265mm
Weight About 5.8kg
Expand more!
Air Fluoride Sampling DeviceAir Fluoride Sampling Device

Click on the inquiry